1) This deviation is caused by equipment and facilities. Because the main peak of the chromatographic column is bifurcated, the experimental results cannot be calculated, the map data is invalid, and the original sample package needs to be re-tested. This deviation has no effect on the product quality evaluation. 2) Check the A0727# chromatographic column in the past two years, only once tested the related substance items of sterile imipenem foreign registration standards, and the results normally met the quality requirements; at the same time, the other products tested did not show the abnormality of the main peak bifurcation, the system is applicable The performance meets the requirements, so the experimental results detected by the chromatographic column in the early stage are valid and risk-free. 3) The chromatographic column number A0727# used for this deviation is no longer suitable for the detection of sterile imipenem foreign registered standard related substances, and the column usage record and its cover note that the column is not suitable for sterile imipenem Foreign registration method related substance item testing. 4) In order to avoid recurrence of similar incidents, the training of liquid-phase personnel using this deviation as a case emphasizes: The detection of related substances in the foreign registration method of sterile imipenem requires special chromatographic columns and cannot be mixed with other products.