At 10:38 p.m. on the 2019.10.26th, Cai Haibing was mid-way with the elution of the 1020-B191004 batch of ion exchange elutions, and found that the thermometer of the R2201a tank could not show the temperature. The process requires a medium-temperature control of 5.0 to 15.0 degrees C. <br><br>The most likely cause of this deviation is that the medium-to-tank R2201a thermometer B-TT02-Y33-15 battery is soldered on the board and cannot be replaced directly. The thermometer battery is a consumable, in the long-term use of the battery will slowly drain, resulting in the temperature can not be displayed. <br><br>When the deviation is found, the medium-sum tank R2201a is used to store Bolamycin extraction 1020-B191004 batch T2201a, b-ion junction column elution, ready to be moderated. Bolemycin process procedures require a temperature control of 5.0 to 15.0 degrees.<br>The metering department confirmed that the Battery of the B-TT02-Y33-15 digital thermometer was empty and could not be mixed directly. In order not to affect production, the same model of the same specifications of R2201b tank thermometer B-TT02-Y33-14, battery power sufficient, 2018.04.28 to check, valid until 2020.04. 27, within the validity period of the school test, the temperature verification range covers the use of 5.0 to 15.0 degrees C range, replacing the B-TT02-Y33-15 digital thermometer for measuring the medium and tank R2201a's medium-liquid temperature.<br>The thermometer probe is inserted into the tube of the glass reaction tank thermometer, does not come into direct contact with the tank infusion, has a small impact on the elution and fluid quality of bolemycin ion exchange, while follow-up follow-up observation of the Bolamycin extraction batch (1020-B191004) and liquid yield and bolrepmycin A2, B2 gel elution liquid, to see if the purity meets the requirements.<br><br>1. Remove the B-TT02-Y33-15 thermometer of the tank R2201a, which is depleted.<br>2. Replace the R2201b can B-TT02-Y33-14 thermometer with the R2201a tank.<br>3. Replace the rinted r2201a digital thermometer with a thermometer that can replace the battery.<br>4. Track and observe the bolamycin extraction batch (1020-B191004) and the liquid yield and Bolamycin A2, B2 gel elution, to see if the purity meets the requirements.<br>5. Check whether the other equipment of the Boralymycin extraction post (CH-490) is not convenient to replace the battery thermometer, depending on the results of the inspection to replace.
正在翻译中..