The root cause of this deviation is that a wire in the connecting cable plug between the temperature and humidity recorder and the printer is welded with the plug when leaving the factory, resulting in accidental disconnection during use, resulting in disconnection of the printer. After preliminary investigation by the company's metrology personnel, the fault cannot be eliminated on site, so the temperature and humidity recorder and printer were removed and taken back to the special maintenance room of Peinan Class in Y70 Building for inspection. During the inspection period (March 8, 2021-March 12, 2021), the temporary room temperature humidity of the intermediate was recorded with high frequency (2 hours/time); After the repair, install and reprint the temperature and humidity curves that were not printed successfully from 2021.03.01/00: 00: 00-2021.03.07/23: 59: 59. The intermediate temporary storage room is at room temperature, and all the materials stored in it are stored at room temperature. The temperature and humidity recorder can record the temperature and humidity normally, but can not print the temperature and humidity curve normally, which will not affect the quality of materials. The materials can undergo the next reaction without affecting the quality of products.
正在翻译中..