This deviation is mainly due to the method and has nothing to do with the quality of the product itself. The main peak area of cyclization reaction solution in batch 3046011N210106-02 is too small to judge the content of TCG1 in sample solution. However, when processing the atlas, analyst Hu Qinxiang immediately recognized that the peak area was too small, and timely notified IPC supervisor to report the deviation. The preliminary analysis of the deviation was due to the uneven liquid removal during the preparation process of the analyst. Because the intermediate is special, the sample cannot be stored for a long time, and the results cannot be obtained after the deviation investigation is finished. Therefore, the sample is retested as it is, and the retest results are qualified, while the original atlas and data are invalid and retained, and a report is issued based on the retest results. Investigating the abnormal events and deviations in the laboratory in the past, Hu Qinxiang has a low overall error rate since taking up his post, and has been careful in his work. In the past three months, there have been no cases similar to deviations. In addition, the analyst has tested 7 batches of cyclization reaction solutions in history, which can be regarded as relatively close and has no influence on the historical test results. The reason of this deviation has nothing to do with the quality of the product itself, so it does not affect the subsequent intermediate quality inspection. See D-TZ-API-IP-21001-R for details.