The equipment was operated according to the standard operating procedures for blister packaging of pre flushing syringe syringe. Three groups of parameters including maximum, minimum and intermediate value of process parameters in the existing process card were selected, and each group of parameters was taken as a test group for sample test. First, set the selected parameters, preheat the equipment to the target temperature, maintain for 5-10 minutes until the temperature is completely stable, and then conduct the molding test of the film. During the test, the technical department shall record the test data, the quality department shall take samples and inspect the appearance of blister for each group of test samples, and fill in the attached table 2 "blister packaging process verification - blister preform appearance inspection record", and Annex 3 "blister packaging process verification - blister forming appearance inspection record".<br>