This deviation is caused by the lack of detailed documents and negligence of personnel. After correction on January 19, 2020, it was re-checked, and the result was qualified, which had no risk to product inspection. The last cleaning time of atomizer and burner of this equipment is December 23rd, 2020. It is necessary to review the samples detected by flame method after this time, and the results show that there is no risk for sample detection. In order to better guide the daily calibration operation of metrological personnel, it is planned to upgrade the program H0-TM-21336 "Calibration Procedure for PINAACLE 900T Atomic Absorption Spectrometer", add relevant regulations such as cleaning and maintenance of instruments before annual calibration, and train personnel. In order to better maintain the instrument, it is planned to upgrade the program H0-TM-22132 "Operation and Maintenance Procedures of PINAACLE 900T Atomic Absorption Spectrometer", refine the cleaning methods of the combustion head, increase the cleaning methods and cleaning cycles of the atomizer, and train relevant personnel.