At 10: 25 on January 20, 2021, analyst Chen Ping processed the 3046011N210105-01 batch of samples of tigrel alkylation reaction solution on the second floor laboratoryNo. A-CAY-Y66-11 of Building Y66, and found that the main peak area (198474) was small, far lower than the historical detection level, and the historical detection batch (such as 304600The root cause of this deviation is that the staff used the sampler for sampling operation for the first time, but did not practice it for a long time after training. During sampling, the feed liquid pressed into the sampler for the first time was not collected into the beaker, but was directly sampled as the target sample, which led to the diluted reaction liquid sample, resulting in deviation, which was due to personnel reasons.After re-sampling and testing of the Tigrinol alkylation reaction solution in 3046011N210105-01, the test results meet the requirements, and the peak area of the main peak is consistent with the historical batch trend, without risk. Looking back at the historical production batches, no similar deviation events occurred, which had no impact on the product quality of the historical batches. In order to prevent similar incidents from happening in the future, relevant personnel were retrained in H3-SOP-61042 "Intermediate Sampling Operation Procedures", emphasizing that the feed liquid pressed into the sampler for the first time should be collected and processed before sampling the reaction liquid of Tigrinol alkylation.CP-TZ-API-D-TZ-API-IP-21005-21001: 3046011N210105-01 batch of tigrel alkylation reaction solution was re-sampled for detection.CP-TZ-API-D-TZ-API-IP-21005-21002: Re-train relevant personnel in H3-SOP-61042 Intermediate Sampling Operation Procedures, emphasizing the need to collect and treat the feed liquid pressed into the sampler for the first time before sampling the tigrillo alkylation reaction liquid.
正在翻译中..