At 13: 48 on January 14, 2021, Duan Guangyu, Department of Analysis Technology, rechecked the annual calibration data of atomic absorption spectrometer (MetrologyNo.: A-CA-ZKZX-2008) in room WS05 5422, and found that the detection limit of flame method was 0.051ug/ml, which did not meet the acceptance standard (≤0.02ug/ml).The root cause of this deviation is that the SOP document is not detailed enough in terms of equipment cleaning, and the verifier Chen Demin did not pay attention to the need to clean the atomizer and combustion head before calibration. It is caused by insufficient details of documents and negligence of personnel.