Confirmation method: Positive and negative pressure sealing test is carried out on the sample sample respectively. Positive pressure: Apply axial force of 30N to the push bar and hold for 30 seconds to test the syringe cap area, the plug area for liquid leakage. Negative pressure: Fill the container used in the test with meditate test fluid, place the sample in the container, and completely immerse the entire sample in the test fluid. Place the container containing the sample in a vacuum box, vacuum the vacuum box to 35kPa±5kPa for 30 minutes, return to atmospheric pressure for another 30 minutes, then remove the sample from the container, rinse the sample with purified water, wipe it clean with absorbent paper, visually check the syringe for dye entry.<br>Acceptable standard: the syringe should not leak when positive pressure test, negative pressure test, no dye should enter the inside of the syringe.